…Classic, and a release gate Hardware-in-the-loop tests of ShimmerDriver/ShimmerDriverPC on a PC with paired units, through ShimmerPC with its timers on, as Consensys and MATLAB run it. - JavaHilScenariosTest (hilTest task, skipped without -Phil.port, left out of the normal test task): streamRest, midStreamSet, reconnectLoop and inStream. inStream requests each in-stream reply (status, directory, battery) and then overlaps SET_RWC with status and battery requests, the DEV-1262 case. Every API call runs under a watchdog that dumps every thread with its locks. - HilShimmerPC, a test-only subclass, counts in-stream replies by type, decoded statuses and battery readings, and captures every driver log line, without changing what the driver does. The configure step does not write the calibration dump back, so runs leave the units' stored calibration alone. - StreamAnalysis is the Android HIL's class (same rules on both); its tests run in CI. - tools/hil/run-java-hil.ps1 finds each unit's COM port from its MAC, builds once, runs each scenario in its own Gradle run under a time limit, checks every report came back, and with -Report posts the commit status java-hil (full default run, clean pushed commit only). - Release Java API stops unless java-hil passed on the commit, or hil_waiver gives a Jira key; the release notes record which. Co-Authored-By: Claude Opus 5.5 <[email protected]>
There was a problem hiding this comment.
Copilot was unable to review this pull request because the user who requested the review has reached their quota limit.
An organization owner can choose the repository organization as the billing source for reviews requested by members with a paid Copilot license. Learn about Copilot code review billing. Review the Copilot code review billing settings.
…ink Windows can't open midStreamSet and inStream now start SD logging as Consensys does (DEV-436): stop streaming, start logging on the quiet link, then stream again. SD logging is still stopped mid-stream. Consensys never writes the RTC mid-stream, so the scenarios no longer do either; that is what caused DEV-1245's out-of-order timestamps on the Shimmer3R. The stream is judged per stretch between the deliberate restarts. Loss within 1 s of an SD logging stop is reported but does not fail, because the unit pauses while it closes its file. Loss near a battery read fails, and loss anywhere else is held to the limit. A connect whose Bluetooth link can't be opened (CONNECTION_FAILED; jssc reports "Port busy") is tried once more, and connectRetries in the report counts those retries. Co-Authored-By: Claude Opus 5.5 <[email protected]>
There was a problem hiding this comment.
Copilot was unable to review this pull request because the user who requested the review has reached their quota limit.
An organization owner can choose the repository organization as the billing source for reviews requested by members with a paid Copilot license. Learn about Copilot code review billing. Review the Copilot code review billing settings.
The pattern expected the command as one word, but the driver logs it as "[0x72] GET_STATUS_COMMAND". So every commandTimeouts count was 0, and the "no command timeouts" checks could never fail. Checked against the bench logs so far: it now finds all 14 timeouts in them. Co-Authored-By: Claude Opus 5.5 <[email protected]>
There was a problem hiding this comment.
Copilot was unable to review this pull request because the user who requested the review has reached their quota limit.
An organization owner can choose the repository organization as the billing source for reviews requested by members with a paid Copilot license. Learn about Copilot code review billing. Review the Copilot code review billing settings.
The accel check was skipped when the mean was NaN, so a stream with no calibrated accel passed. A Shimmer3 (3E36, LogAndStream v1.1.7) did exactly that: all 7655 samples NaN after the driver read its calibration dump. Co-Authored-By: Claude Opus 5.5 <[email protected]>
There was a problem hiding this comment.
Copilot was unable to review this pull request because the user who requested the review has reached their quota limit.
An organization owner can choose the repository organization as the billing source for reviews requested by members with a paid Copilot license. Learn about Copilot code review billing. Review the Copilot code review billing settings.
…trying it Every "Port busy" the HIL ever met was another program holding the COM port, never a flaky first connect. Two HILs shared a unit, and that sharing produced the misleading DEV-1276. A retry only lets the run go on while the unit is shared, so its results mean nothing. So a connect is no longer retried. When it fails, the report says why: - the port is open in another program (jssc "Port busy"), which the HIL now sees because HilShimmerPC records the messages the driver prints with consolePrintException; - the link could not be opened for another reason; - the unit did not initialise in time. The README now says that nothing else may use the units during a run, and why a BLE host is the dangerous case. Co-Authored-By: Claude Opus 5.5 <[email protected]>
While a unit streams, four app threads call readBattery() and readStatusLogAndStream() and check the queue length, in tight loops for 60 s, keeping the queue under 8. Each of those calls locks the device and then the instruction list, so one is nearly always under way while the IO thread, holding the list, fetches the next instruction. That is the window DEV-1266's lock inversion needs. No SD logging is involved. The scenario fails if the calls and replies stop for 10 s, or if the JVM finds a deadlock. Either way the events get every thread's stack, and the unit is left alone. On a Shimmer3R (v1.1.21) without the DEV-1266 fix (#334), it deadlocked within 0.3 s in both runs. With the fix, two 60-second runs were clean, with about 670 calls and 0 timeouts each. It runs by default, so the release gate includes it. Until #334 merges, the gate fails on a deadlock. Co-Authored-By: Claude Opus 5.5 <[email protected]>
A Shimmer3 (LogAndStream v1.1.7) sometimes gives up one sample slot to answer GET_VBATT mid-stream: packet n, the battery reply, then packet n+2. That was 4 skips in 33 battery reads. It's the unit, not the host: there were no parser errors, and builds with and without the driver fixes skip at the same rate. Filed for the firmware as DEV-1297. The check charged it as packet loss near a battery read, so midStreamSet failed on the Shimmer3 about one run in three. A gap near a battery read is now excused only when it is exactly one sample and the packet after it arrived within 25 ms of a decoded battery reply. Such a gap is reported, as skippedAtBatteryReplies plus an event naming DEV-1297. Anything else near a battery read still fails. A packet the driver dropped would leave stray bytes, which checkDriverLog() fails on anyway. HilShimmerPC now records when each battery reply is decoded. Checked on 3E36: 4 midStreamSet runs all passed. Two of them reported one skip at a battery reply, and there was no other loss near any read. Co-Authored-By: Claude Opus 5.5 <[email protected]>
What this adds
Hardware-in-the-loop tests for the Java driver. A PC connects to real Shimmer3 and Shimmer3R units over their paired Bluetooth Classic serial ports, through
ShimmerPCwith the driver's timers on, as Consensys and MATLAB run it. One command builds, runs every scenario on every unit, and prints pass/fail:Setup, scenarios and output are in
tools/hil/README.md.JavaHilScenariosTest, run by a newhilTesttask. Without-Phil.portit is skipped, and the normaltesttask leaves it out, so CI is unaffected.streamRest: stream at rest; check loss, timestamps, rate, silence and accel.midStreamSet: SD logging started the way Consensys does (stop streaming, start logging on the quiet link, stream again; DEV-436), stopped mid-stream, and battery reads while streaming.reconnectLoop: five connects on oneShimmerPC.inStream: each in-stream reply (status, directory, battery) on its own. Then the DEV-1262 case: with the unit streaming and logging (started the Consensys way), stop SD logging mid-stream and request status and battery at once.queueStress: while streaming, four app threads queue reads and check the queue length in tight loops for 60 s, under a hang watchdog. That is DEV-1266's deadlock on demand, with no SD logging (DEV-1266: Take the instruction list's two locks in one order #334).Stream checks per stretch between the deliberate restarts. Loss within 1 s of an SD logging stop is reported but doesn't fail, because the unit pauses while it closes its file. Loss near a battery read fails, and loss anywhere else is held to the limit.
No connect retries, and a failed connect says why. "Port busy" means another program on the PC has the COM port open: another HIL, Consensys, a serial terminal. That fails the scenario at once, naming the cause, because results from a unit shared with another host mean nothing. A
CONNECTION_FAILEDfor any other reason, or a unit that never initialises, is reported as such. The README says nothing else may use the units during a run, and explains why a BLE host is the dangerous case: a Shimmer3R accepts it during a Classic stream.Watchdog: every API call runs under one. If a call hangs, the report gets every thread's stack, with the locks each holds and any deadlock the JVM detects (DEV-1266).
HilShimmerPC, a test-only subclass, does the counting without changing what the driver does:Its configure step does not write the calibration dump back (
configureFromClonedoes by default, AA-246), so runs leave the bench units' calibration alone.StreamAnalysisis the Android HIL's class, so both HILs judge a stream by the same rules.StreamAnalysisTestruns in CI.tools/hil/run-java-hil.ps1:-Reportposts the commit statusjava-hil, for the full default run on a clean, pushed commit only.Release Java API now stops unless
java-hilpassed on the commit, orhil_waivergives a Jira key. The release notes record which, and dry runs only warn.Heads-up: the gate needs the three driver fixes this PR found
On master as it is, the gate fails:
inStreamon DEV-1291: in-stream replies that arrive with no data packet before them are thrown away. Fixed in DEV-1291: Take an in-stream ACK that comes with no data packet before it #332.streamReston DEV-1275: the unit is reported ready early, and a Shimmer3's accel comes out NaN. Fixed in DEV-1275: Report a Shimmer ready only once its connect sequence has ended #333.queueStresson DEV-1266: the deadlock. Fixed in DEV-1266: Take the instruction list's two locks in one order #334.With all three on top of this branch, a full run passed on both units. Until they merge, a release needs
hil_waiver.DEV-1245 no longer trips the HIL. That fault came from writing the RTC mid-stream, and Consensys never does that. The scenarios now follow Consensys's order, so they don't either.
Tested on the bench (2026-10-21)
A full run of all four scenarios, on the PR's head, whose driver is master's (
dffb582c) unchanged. Bench PC, units paired over Classic, the driver's default CRC (1 byte), the Shimmer3 undocked, and nothing else using either unit.Every connect succeeded first time, and no packet was lost near a battery read or anywhere away from an SD logging stop.
Earlier runs today overlapped two other HILs (C# and shimmer-logandstream) on the same units. Every "Port busy" this PR saw came from that clash, and so did DEV-1276's evidence. Those runs aren't counted here.
Checked after removing the retry (2026-10-22, Shimmer3 3E36):
streamRestfails at once with "COM43 is open in another program (another HIL, Consensys, a serial terminal?): jssc reports "Port busy"".reconnectLoopconnected all 5 cycles. Two other faults showed up on the way, neither from this change. Both runs (11:48 and 11:50) also overlapped a C# HIL streaming another unit through the same PC Bluetooth adapter, which it measured at about 240 ms radio silences, so treat them as seen under a busy radio:GET_SAMPLING_RATEgot its reply 273 ms late, which was read asGET_SHIMMER_VERSION's ACK. The version was never read, and the connect timer gave up after 20 s ("Lost connection").What it found:
midStreamSetfailed on the Shimmer3 about one run in three. Afterwards 4 of 4 runs passed, two of them reporting one skip.queueStresson a Shimmer3R deadlocked within 0.3 s in both runs without DEV-1266: Take the instruction list's two locks in one order #334: the IO thread held the instruction list ingetInstruction()and waited for the device, while a stress thread held the device and waited for the list. With DEV-1266: Take the instruction list's two locks in one order #334, two 60-second runs were clean, with about 670 calls and 0 timeouts each.STOP_LOGGING_ONLYthe Shimmer3R stops sending data packets for about 100 ms while it closes its file. A status request answered in that pause sendsFF 8A 71 ..straight after the previous ACK, with no data packet before it.processPacket()only looks for an ACK after a data packet (there is aTODO: ACK in bufferTemp[0] not handled), so it throws the reply away. The command then times out after 2 s, and the timeout'sclearAllInstructions()silently drops the battery request queued behind it. Android's Classic parser (Shimmer.java) has the sameTODO.streamRest, a second symptom of DEV-1275. The driver reported the unit ready while its connect sequence was still reading the calibration dump. The HIL's configure then took over the queue mid-transfer, and the dump's second chunk was cleared as junk. The driver kept 128 of the dump's 536 bytes, and all 7655 samples were NaN. The dump itself is fine: its accel block holds the standard defaults. In the other scenarios the configure came early enough that no dump was read, and accel read about 10.9 m/s² at rest from this unit's stored calibration (11.1 on v1.1.5 this morning), so that calibration is off too. None of this code has changed since 0.11.9_beta.[0x72] GET_STATUS_COMMANDform, socommandTimeoutswas always 0. Fixed; it now finds all 14 timeouts in the bench logs.Also checked:
gradlew :test:StreamAnalysisTest8 of 8; the hardware scenarios are left out.gradlew :hilTestwithout a port: all four skipped.Notes
Shimmer-Java-Android-API, because ShimmerDriverPC'ssettings.gradlefinds its siblings that way. In a worktree, use a junction; the README explains how.hilTestshould be run with--no-watch-fs(the runner does). Gradle's file watcher misses edits made through a junction's real path, and would run stale classes.🤖 Generated with Claude Code